<font face="Times New Roman" color="#000000" size="3"> </font><font color="#000000">Hydrogen sulfide (H2S) is produced as a result from the degradation of biomass. Far flowing distances, long flowing times as well as saving water at constant pollution loads cause the loads to start decomposition even within the channel. </font>
<font color="#000000">This in turn leads to heavy corrosion of concrete channel constructions and, in combination with water, destroys the channel constructions in the long run. To avoid damaged channels it makes sense to investigate the hydrogen sulfide concentration, which allows to take the appropriate corrective measures.<br></font>
<font face="Times New Roman" size="3"> </font>
<font face="Times New Roman" color="#000000" size="3"> </font>
NivuLog H2S Bluetooth | |
---|---|
Transmission | Bluetooth Smart |
Display | 1,5" colour OLED display |
Battery life | Sensor module 18 months at 1 min. measurement cycle, Gateway 36 months |
Protection | IP 66 |
Pressure range | atmospheric ±10 % |
Operating temperature | -10 °C ... +40 °C Relative humidity: 15 % ... 90 % r.F. |
Sensor modules | H2S Gas Nominal: 0 ... 200 ppm, max. overload: 1000 ppm Resolution: 0.25 ppm, repeat accuracy: 1 %m T90, time: 35 sec., wear: 0 ... 100 % |
Measurement range | Internal Temperature: -20 °C ... +50 °C Internal relative humidity: 0 ... 100 % r.H. State of battery charge: 0 ... 100 % Bluetooth / GSM signal quality: -50 dBm ... 110 dBm |
Readings memory | approx. 1380 days with 10 min. recording cycle |
Enclosure | 166 x 106 x 61 mm (H x B x T) |
Material | Noryl GTX 973 / PC |
Specifications subject to change.
NivuLog H2S Bluetooth GSM | |
---|---|
Transmission | Bluetooth Smart 2G GPRS 900 MHz / 1800 Mhz 3G WCDMA 900 MHz / 2100 Mhz |
Display | 1,5" colour OLED display |
Battery life | Sensor module 18 months at 1 min. measurement cycle, Gateway 24 months at 4 h transmission interval |
Protection | IP 66 |
Pressure range | atmospheric ±10 % |
Operating temperature | -10 °C ... +40 °C Relative humidity: 15 % ... 90 % r.H. |
Sensor modules | H2S Gas Nominal: 0 ... 200 ppm, max. overload: 1000 ppm Resolution: 0.25 ppm, repeat accuracy: 1 %m T90, time: 35 sec., wear: 0 ... 100 % |
Measurement range | Internal Temperature: -20 °C ... +50 °C Internal relative humidity: 0 ... 100 % r.H. State of battery charge: 0 ... 100 % Bluetooth / GSM signal quality: -50 dBm ... 110 dBm |
Readings memory | approx. 1380 days with 10 min. recording cycle |
Enclosure | 166 x 106 x 61 mm (H x B x T) |
Material | Noryl GTX 973 / PC |
Specifications subject to change.
Compact, portable measurement instrument with readings memory for data collection of H2S concentration in channel networks