Flow Sensors
CSP Correlation Wedge Sensor
Part No. CSPLarge measurement range for high flow levels
Flow velocity sensor for continuous measurement in slightly to heavily polluted media in channels with mainly high filling levels.
- For flow measurements with very high accuracy through ultrasonic cross-correlation method
- Absolutely stable zero point and drift-free
- Measurement in slightly to heavily contaminated media
- Simple installation without additional fixtures and optimised mounting accessories
- Installation possible under process conditions
The CSP flow velocity sensors are suitable for continuous measurement in slightly to heavily polluted media in channels with mainly high filling levels. The wedge sensors are designed for mounting on the channel floor or on the channel wall. In combination with the NivuFlow Mobile 750 transmitter, they make a complete flow measurement system..
Typical applications for the CSP wedge sensor
- Measurement campaigns in the sewer network
- Precipitation discharge measurements
- Wastewater treatment plant: inlet, outlet
- Industrial wastewater networks
- Large collecting sewers
- Particularly suitable for large sewers with fill levels from 500 mm
| Allgemein | |
|---|---|
| Measurement principle |
|
| Measurement frequency | 1 MHz |
| Protection | IP 68 |
| Ex-Approval |
|
| Operating temperature | −40 °C to +80 °C |
| Storage temperature | −30 °C to +70 °C |
| Operating pressure | max. 4 bar (combi sensor with pressure cell max. 1 bar) |
| Cable lengths | 7 / 15 / 20 m |
| Cable types | Combi sensor with/without pressure measurement: LiYC11Y 2x(2x AWG24/7-CAT7) + PA 1,5/2,5 + (4x AWG26/7) |
| Outside cable diameter | Combi sensor with/without pressure measurement: 9,7 mm ±0,2 mm |
| Sensor types |
|
| Construction | Wedge sensor for mounting on the channel floor or the channel wall |
| Medium-contacting materials | Polyurethane, stainless steel 1.4571, PPO GF30, PEEK, PA6 |
Flow velocity measurement |
|
| Measurement range | −350 cm/s to +600 cm/s |
| Quantity of scan layers recorded and computed | max. 32 |
| Zero point drift | absolutely stable zero point |
| Sensor connection |
|
| Error limits | < 1 % of measurement value (v > 1 m/s) |
| Beam angle | ±5 degrees |
Level measurement water-ultrasonic sensor |
|
| Measurement range | 0.08 to 500 cm |
| Zero point drift | Absolutely stable zero point |
| Measurement uncertainty | +/− 2.0 mm or 0.2% of measured range (whichever is greater) |
Level measurement – Pressure |
|
| Measurement range | 0 to 500 cm |
| Zero point drift | max. 0.75 % of final value (0 - 50 °C) |
| Measurement uncertainty | standing medium: +/− 0.5 of final value (within a range of 0.05 – 5 m) |
Temperature measurement |
|
| Measurement range | −20 °C to +50 °C |
| Measurement uncertainty | ±0.5 K |